Стандартный образец для настройки дефектоскопа на фазированных решетках, сталь 1018
Introduction
Phased Array Calibration Standard Blocks (PACS) are used for angle beam verification, probe angle exit points, calibration for wedge delay, sensitivity and DAC/TCG for thicknesses up to 2".
Inch Version
The three radii (.500", 1.000" and 2.000") allow for exit point verification, velocity and sound path calculations. This block contains five holes at 3/64" diameter drilled through the 1.000" width, located at .100", .200", .400", .600", .800", 1.200", 1.400", 1.600", 1.800" and 1.900" from the respective scanning surface.
Inch Dimensions: 18.0" x 2.0" x 1.0"
Weight: 6.5kg
Стандартный образец для настройки дефектоскопа на фазированных решетках, сталь 1018
<div style="text-align: center;">
Introduction
</div>
<div style="text-align: center;">
</div>
<br>
Phased Array Calibration Standard Blocks (PACS) are used for angle beam verification, probe angle exit points, calibration for wedge delay, sensitivity and DAC/TCG for thicknesses up to 2".<br>
<br>
Inch Version<br>
The three radii (.500", 1.000" and 2.000") allow for exit point verification, velocity and sound path calculations. This block contains five holes at 3/64" diameter drilled through the 1.000" width, located at .100", .200", .400", .600", .800", 1.200", 1.400", 1.600", 1.800" and 1.900" from the respective scanning surface.<br>
<br>
Inch Dimensions: 18.0" x 2.0" x 1.0"<br>
Weight: 6.5kg <br>